Introduction
Recent advances in medical logistics have highlighted the critical need for rapid and reliable transport of biological specimens, especially in geographically remote or hard-to-access regions [1-3]. Blood plasma samples used for coagulation analysis are among the most sensitive biomaterials, requiring strict adherence to pre-analytical protocols – temperature control, processing time, and storage conditions – to preserve diagnostic integrity.
The Thrombodynamics assay, a sophisticated global coagulation test, exemplifies these stringent requirements [4-5]. By simulating vascular‑wall injury and modeling the spatial dynamics of thrombus formation, it offers exceptional sensitivity to anticoagulant therapy and can effectively detect both thrombotic and bleeding tendencies. Given this high sensitivity, maintaining sample integrity during transport is essential for reliable results [6-7].
Traditionally, plasma stability has been limited to several hours post-collection, which markedly restricts diagnostic capabilities in areas lacking immediate laboratory infrastructure. Ground transportation often subjects samples to prolonged vibration, mechanical shocks, and temperature fluctuations, all of which can degrade sample quality. While air transport is faster, it is frequently cost-prohibitive and logistically demanding, particularly in sparsely populated or inaccessible regions [8-10].
Unmanned aerial vehicles, commonly known as drones, have emerged as a novel and potentially transformative solution for biological sample logistics. Several studies have demonstrated the feasibility of UAV-based transport for whole blood and other biomaterials (11-15). However, the impact of UAV transportation on the results of highly sensitive assays, such as Thrombodynamics, remains unexplored.
This study aims to fill that knowledge gap by evaluating the effects of UAV-based transportation on key Thrombodynamics parameters and establishing operational conditions that preserve plasma sample integrity during drone delivery. To achieve this, we employed the InnoVtol-3s VTOL UAV – a hybrid vertical-take-off and landing platform capable of covering distances up to 250 km with low mechanical load and tightly controlled temperature conditions.
Material and Methods
Study Design and Population
This prospective study enrolled 30 healthy adult volunteers (11 men and 19 women), aged 22-38 years (mean=30±8 years). All participants provided written informed consent before inclusion. The study adhered to national ethical regulations, the principles of Good Clinical Practice (GCP) of the Russian Federation, and the Declaration of Helsinki. Ethical approval was obtained from the Ethics Committee of the Pacific State Medical University, Ministry of Health of the Russian Federation.
Blood Sample Collection and Plasma Preparation
Venous blood was drawn from each participant into three 3.6-mL vacuum tubes containing 3.2 % sodium citrate (0.106 mol/L). Platelet-free plasma was prepared according to standard pre-analytical laboratory protocols, including double centrifugation, and stored under controlled conditions before testing.
Experimental Groups
Plasma samples were divided into three groups:
- Reference group (control): Samples stored at room temperature (22 °C) without transportation.
- UAV transportation group: Samples transported by the InnoVtol-3s VTOL UAV over a distance of up to 250 km.
- Immediate analysis group: Samples analyzed immediately after blood collection without storage or transport.
Thrombodynamics Assay Procedure
Thrombodynamics testing was performed using the Thrombodynamics Analyzer T‑2 system (HemaCore LLC, Moscow, Russia) with the appropriate reagent kits. Clot formation was initiated by contact of plasma with a surface-immobilized activator, and the following parameters were assessed:
- Lag time (Tlag, min): Delay between activator contact and clot initiation.
- Clot growth rate (V, µm min⁻¹): Average clot growth rate, calculated over an interval of 15-25 min after the start of growth.
- Initial clot growth velocity (Vi, µm min⁻¹): Average growth rate between 2-6 min.
- Stationary clot growth velocity (Vst, µm min⁻¹): Average growth rate between 15-25 min.
- Clot size (CS, µm): Maximum clot size at 30 min.
- Clot density (D, arbitrary units): Optical measurement reflecting fibrin network density.
- Spontaneous clotting time (Tsp, min): Time to appearance of spontaneous clots in plasma not contacting the activator.
UAV Transportation Conditions
The InnoVtol-3s VTOL UAV was used to transport the plasma specimens. This hybrid UAV combines an internal-combustion engine for horizontal flight with electric motors for vertical take-off and landing, thereby eliminating the need for launch devices. The drone is equipped with an advanced temperature-regulation system and vibration-dampening mechanisms. Temperature and vibration levels were continuously monitored throughout transport.
Statistical Analysis
Statistical analysis was performed using StatTech v. 4.7.3 (StatTech LLC, Russia). Quantitative indicators were evaluated for normality with the Shapiro-Wilk test. For normally distributed indicators, arithmetic means (M) and standard deviations (SD) described the sample distribution, and the 95% confidence interval (95% CI) was provided as a measure of representativeness for the mean values. For non-normally distributed data, the median (Me), lower quartile (Q1), and upper quartile (Q3) were used for description.
A comparison of three or more groups based on a quantitative indicator was performed using one-factor analysis of variance. Posterior comparisons were conducted with Tukey’s test (assuming equal variances). If the Kruskal-Wallis test did not reveal significant differences, further pairwise comparisons were not performed, as the general test had already determined that no significant differences were observed between groups. If significant differences were found, we performed pairwise comparisons using the Dunn test with Holm correction, considering differences significant at p<0.05. Without correcting the significance level, the probability of obtaining false‑positive results increases; for example, by applying the Bonferroni or Benjamini-Hochberg corrections.
Results
UAV Transportation Conditions
To reduce environmental effects on plasma specimens during flight, the InnoVtol-3s VTOL UAV was fitted with an advanced temperature-control system and vibration-dampening mechanisms. Temperature and vibration parameters were continuously recorded throughout transport (Figures 1-2). Flight durations stayed within the time limits recommended by the Thrombodynamics assay manufacturer, thereby preserving plasma specimen stability.
Figure 1. Vibration level in the InnoVtol-3s UAV during test flights.
Figure 2. Temperature fluctuations during plasma transportation.
The InnoVtol-3s is a hybrid UAV that uses an internal-combustion engine for horizontal flight, giving it a longer range than multi-rotor UAVs. Although this design increases vibration during operation, tests have shown that all control and measurement systems remain within acceptable limits. Using the electric version of the UAV, which produces less vibration, would likely improve the stability of the bioassay container. Consequently, multiple UAV models can be employed to transport blood plasma for thrombodynamic analysis, while additional data could be collected by expanding the UAV fleet and increasing specimen volume.
Thrombodynamics Parameter Analysis
Analysis of the key Thrombodynamics parameters revealed no statistically significant differences among the three experimental groups (immediate analysis, reference, and UAV-transported specimens). As summarized in Table 1, clot growth velocities (Vi, Vst), lag time (Tlag), clot size (CS), and clot density (D) remained consistent across all groups. Statistical analysis confirmed that the variations were within acceptable analytical ranges (coefficient of variation <4%). Notably, we observed no spontaneous clot‑formation events (Tsp) in any sample, indicating the absence of pre‑analytical coagulation activation during UAV transport. This parameter is evaluated by the system as a binary presence/absence and is therefore not represented as a numeric value.
Table 1. Growth-rate and related parameters by sample type
|
Parameter |
Sample type |
Mean±SD |
95% CI |
n |
p-value |
|
Clot growth rate (µm min⁻¹) |
UAV-derived |
29.2±2.8 |
27.9-30.5 |
30 |
0.388 |
|
Lab‑stored reference |
30.4±2.9 |
29.0-31.7 |
30 |
||
|
On‑site analysis |
29.4±2.8 |
28.1-30.7 |
30 |
||
|
Lag time (min) |
UAV-derived |
0.9±0.1 |
0.8-0.9 |
30 |
0.737 |
|
Lab‑stored reference |
0.9±0.1 |
0.8-1.0 |
30 |
||
|
On‑site analysis |
0.9±0.1 |
0.8-1.0 |
30 |
||
|
Initial clot growth rate (µm min⁻¹) |
UAV-derived |
56.9±4.3 |
54.8-58.9 |
30 |
0.779 |
|
Lab‑stored reference |
57.8±4.1 |
55.9-59.7 |
30 |
||
|
On‑site analysis |
57.1±4.5 |
54.9-59.2 |
30 |
||
|
Clot size (µm) |
UAV-derived |
1210.4±95.2 |
1165.8-1255.0 |
30 |
0.511 |
|
Lab‑stored reference |
1244.0±102.9 |
1195.9-1292.2 |
30 |
||
|
On‑site analysis |
1216.4±93.8 |
1172.5-1260.3 |
30 |
||
|
Clot density (AU) |
UAV-derived |
21092.0±1949.6 |
20179.6-22004.5 |
30 |
0.831 |
|
Lab‑stored reference |
21382.0±1923.9 |
20481.6-22282.5 |
30 |
||
|
On‑site analysis |
21430.5±1824.0 |
20576.8-22284.1 |
30 |
Figures 1-2 illustrate vibration levels and temperature fluctuations during UAV flights, while Figures 3-4 present clot density and clot size analyses, further supporting the absence of a significant impact from UAV transport.
Figure 3. Analysis of clot density by sample type.
Figure 4. Analysis of clot area by sample type.
A combined boxplot of standard deviations (Figure 5) confirms that overall variability across all measured parameters remained low and consistent among all groups.
Figure 5. Graph of permissible deviations of the Thrombodynamics test parameters.
To test the hypothesis that there are no significant differences between the two groups, we conducted a paired‑comparison analysis. The results, including relevant statistical indicators are presented in Table 2.
Table 2. Relevant statistical indicators paired‑comparison analysis
|
Parameter |
Sample type |
p-value |
|
Clot growth rate (µm min⁻¹) |
UAV-derived |
0.411 |
|
Exposition |
||
|
Lag time (min) |
UAV-derived |
0.623 |
|
Exposition |
||
|
Initial clot growth rate (µm min⁻¹) |
UAV-derived |
0.674 |
|
Exposition |
||
|
Clot size (µm) |
UAV-derived |
0.321 |
|
Exposition |
||
|
Clot density (AU) |
UAV-derived |
0.904 |
|
Exposition |
Discussion
This study provides the first systematic evidence that transporting plasma specimens via a UAV – specifically the InnoVtol-3s VTOL drone – does not substantially compromise the results of the highly sensitive Thrombodynamics global coagulation assay. The data show that UAV transport preserves plasma specimen quality and diagnostic reliability, as reflected by stable clot growth velocities (Vi, Vst), lag time (Tlag), clot density (D), and clot size (CS). Importantly, the absence of spontaneous clotting indicates that no pre-analytical coagulation activation occurred during transport.
These findings are clinically relevant, especially for overcoming logistical hurdles associated with biological sample transport in remote or hard-to-reach areas [8-10]. Ground transport often entails extended transit durations, mechanical impacts, and thermal instability [11-13]. While air transport offers speed, it is frequently costly and operationally complex. UAV-based delivery therefore emerges as a promising alternative that can surmount these limitations and ensure timely, reliable specimen transfer [12-15].
Prior studies have documented the negative impact of pneumatic tube systems on coagulation assays – including thrombin generation, ROTEM, and TEG – due to excessive mechanical stress [18-19]. In contrast, the InnoVtol-3s drone’s hybrid design, vibration-damping features, and temperature-control systems likely contributed to the absence of such detrimental effects in this investigation.
Our findings support the broader adoption of UAVs in clinical laboratory logistics, offering a reliable method for transporting even highly sensitive specimens. By enabling prompt and dependable diagnostics in underserved or geographically isolated regions, UAV technology may improve clinical outcomes and promote healthcare equity.
Previous research on various transportation methods – such as pneumatic tube systems – has demonstrated measurable effects on coagulation assays, including ROTEM, TEG, and platelet aggregation tests, underscoring the need for rigorous validation of new transport technologies [18, 19]. Consequently, evaluating the potential impact of UAV-based transportation on global coagulation testing is a critical step toward its implementation in clinical laboratory practice.
Future studies should explore alternative UAV platforms, including fully electric VTOL drones, which may further reduce vibration and enhance transport stability. Additionally, assessing UAV performance under a broader range of environmental conditions, over greater distances, and for other sensitive biological materials will be essential.
Conclusion
This study is the first to demonstrate that transporting plasma specimens with a vertical-take-off and landing UAV (InnoVtol-3s VTOL) does not gravely compromise the analytical outcomes of global coagulation testing using the Thrombodynamics assay. All assessed parameters – clot growth velocity, lag time, clot size, and clot density – remained within clinically acceptable limits, and no pre-analytical activation of coagulation was detected.
These findings confirm that VTOL UAVs preserve plasma sample integrity during transit, enabling drone-based delivery for reliable, time-sensitive medical logistics. The approach has the potential to markedly improve access to high-quality diagnostic testing in remote, rural, and logistically challenging regions, thereby accelerating clinical decision-making, enhancing healthcare equity, and improving patient outcomes. Future research should evaluate UAV performance over longer distances and under diverse environmental conditions and explore applicability for other sensitive diagnostic materials.
In summary, VTOL UAV transportation offers a promising and practical solution for integrating advanced diagnostic services into broader healthcare delivery networks.
Limitations
It should be noted that our study was conducted on a relatively small sample size of 30 healthy volunteers, which may limit the ability to assess the stability of certain dynamic parameters, such as the hemostasis system. Consequently, the results cannot be generalized to patients with clotting disorders, those on anticoagulant therapy, or individuals with thrombophilia. These individuals may have more sensitive samples that are prone to external influences, which could potentially skew the results. Therefore, we recommend further research involving larger clinical populations to better understand the effects of these factors on blood clotting and hemostasis.
Author Contributions
Conceptualization: Shumatov V.B., Lebedev S.V., Markelova E.V., Dolgopolov M.S., Devitt D.
Methodology: Dolgopolov M.S., Devitt D.
Formal analysis: Dolgopolov M.S., Devitt D.
Data curation: Dolgopolov M.S., Devitt D.
Writing – original draft preparation: Dolgopolov M.S., Shevchenko O.V.
Writing – review and editing: Markelova E.V., Prosekova E.V.
Supervision: Shumatov V.B., Lebedev S.V., Markelova E.V.
All authors have read and approved the final version of this manuscript.
Funding
This research received no external funding.
Institutional Review Board Statement
The study was conducted in accordance with all relevant national regulations, the ethical principles of the Rules of Clinical Practice in the Russian Federation, and the Declaration of Helsinki of the World Medical Association. It received approval from the Ethics Committee of the Federal State Budgetary Educational Institution of Higher Education, Pacific State Medical University of the Ministry of Health of the Russian Federation.
Informed Consent Statement
The institute’s opt-out registry was consulted to determine whether patients objected to participating in scientific research.
Data Availability Statement
The raw data supporting the conclusions of this article will be made available by the authors upon request and may only be shared anonymously. This restriction is due to hospital policies that prohibit data sharing without clearly defined purposes. Data transfer agreements must be in place before data can be shared.
Conflicts of Interest
The authors declare no conflicts of interest. Funders had no role in the study design, data collection, analyses, interpretation, manuscript writing, or decision to publish.
AI Disclosure Statement
During manuscript preparation, the authors did not use any generative AI or AI‑assisted tools for content generation, data analysis, or figure production.
- Zailani MA, Azma RZ, Aniza I, Rahana AR, Ismail MS, Shahnaz IS, et al. Drone versus ambulance for blood products transportation: an economic evaluation study. BMC Health Serv Res 2021; 21(1): 1308. https://doi.org/10.1186/s12913-021-07321-3.
- Patel MK, El-Khoury JM, Simundic AM, Farnsworth CW, Broell F, Genzen JR, et al. Evolution of blood sample transportation and monitoring technologies. Clin Chem 2021; 67(6): 812-819. https://doi.org/10.1093/clinchem/hvab001.
- CLSI. Handling, transport, processing, and storage of blood specimens for routine laboratory examinations. CLSI guideline PRE04. 1st Ed. Wayne, PA: Clinical and Laboratory Standards Institute. 2023; 96 p. https://clsi.org/shop/standards/pre04.
- Balandina AN, Koltsova EM, Shibeko AM, Kuprash AD, Ataullakhanov FI. Thrombodynamics: a new method to the diagnosis of hemostasis system disorders. Pediatric Hematology/Oncology and Immunopathology 2018; 17(4): 114-126. Russian. https://doi.org/10.24287/1726-1708-2018-17-4-114-126.
- CLSI. Collection, Transport, and Processing of Blood Specimens for Testing Plasma-Based Coagulation Assays. CLSI guideline H21. 6th ed. Clinical and Laboratory Standards Institute, USA. 2024; 86 p. https://clsi.org/shop/standards/h21.
- Startseva O. Quality control of laboratory parameters of the thrombodynamics test in real clinical practice: a cross-sectional study. Russian Clinical Laboratory Diagnostics 2024; 69(7): 56-69. Russian. https://doi.org/10.17816/cld630211.
- Balandina AN, Serebriyskiy II, Poletaev AV, Polokhov DM, Gracheva MA, Koltsova EM, et al. Thrombodynamics – A new global hemostasis assay for heparin monitoring in patients under the anticoagulant treatment. PLoS One 2018; 13(6): e0199900. https://doi.org/10.1371/journal.pone.0199900.
- Panteleev MA, Hemker HC. Global/integral assays in hemostasis diagnostics: promises, successes, problems and prospects. Thromb J 2015; 13(1): 5. https://doi.org/10.1186/s12959-014-0032-y.
- Perlee D, van der Steege KH, den Besten G. The effect of drone transport on the stability of biochemical, coagulation and hematological parameters in healthy individuals. Clin Chem Lab Med 2021; 59(11): 1772-1776. https://doi.org/10.1515/cclm-2021-0513.
- Weekx S, Van Lint P, Jacobs S. The effects of drone transportation on routine laboratory, im-munohematology, flow cytometry and molecular analyses. Clin Chem Lab Med 2025; 63(2): 311-319. https://doi.org/10.1515/cclm-2024-0420.
- Amukele TK, Sokoll LJ, Pepper D, Howard DP, Street J. Can Unmanned Aerial Systems (Drones) Be Used for the Routine Transport of Chemistry, Hematology, and Coagulation Laboratory Specimens? PLoS One 2015; 10(7): e0134020. https://doi.org/10.1371/journal.pone.0134020.
- Johannessen KA, Wear NKS, Toska K, Hansbo M, Berg JP, Fosse E. Pathologic blood samples tolerate exposure to vibration and high turbulence in simulated drone flights, but plasma samples should be centrifuged after flight. IEEE J Transl Eng Health Med 2021; 9: 4000110. https://doi.org/10.1109/jtehm.2021.3053172.
- Sylverken AA, Owusu M, Agbavor B, Kwarteng A, Ayisi-Boateng NK, Ofori P, et al. Using drones to transport suspected COVID-19 samples; experiences from the second largest testing centre in Ghana, West Africa. PLoS One 2022; 17(11): e0277057. https://doi.org/10.1371/journal.pone.0277057.
- Geerlings MJ, Hofste LSM, Kamping EJ, Abdi Z, Tolmeijer SH, Garms LM, et al. Effect of pneumatic tube system transport on cell-free DNA. Clin Chem 2021; 67(2): 434-435. https://doi.org/10.1093/clinchem/hvaa285.
- Aarsand AK, Webster C, Fernandez-Calle P, Jonker N, Diaz-Garzon J, Coskun A, et al. The EFLM Biological Variation Database. https://biologicalvariation.eu.
- Berns AS, Sovetnikov EN, Chebotareva NV, Berns SA, Solonkina AD, Guliaev SV, et al. Evaluation of hemostasis disorders using the thrombodynamic test in patients with chronic glomerulonephritis with nephrotic syndrome. Ter Arkh 2022; 94(6): 738-742. Russian. https://doi.org/10.26442/00403660.2022.06.201558.
- Chebotareva NV, Kharionovskaya EA, Biryukova EA, Berns SA, Vuimo TA. Comparison of thrombodynamic methods and routine hemostasis tests in the evaluation of hypercoagulable syndrome in chronic glomerulonephritis. Ter Arkh 2024; 96(6): 565-570. Russian. https://doi.org/10.26442/00403660.2024.06.202723.
- Pisaryuk AS, Povalyaev NM, Poletaev AV, Shibeko AM. Systems Biology Approach for Personalized Hemostasis Correction. J Pers Med 2022; 12(11): 1903. https://doi.org/10.3390/jpm12111903.
- Minaev AV, Gushchin DK, Kovalev DV, Mwela BM. Hemostasis assessment in Fontan patients using the new thrombodynamics test. Egypt Heart J 2023; 75(1): 39. https://doi.org/10.1186/s43044-023-00365-2.
- Betticher C, Bertaggia Calderara D, Matthey-Guirao E, Gomez FJ, Aliotta A, et al. Global coagulation assays detect an early pro-thrombotic state in children with acute lymphoblastic leukemia. J Thromb Haemost 2024; 22(9): 2482-2494; https://doi.org/10.1016/j.jtha.2024.05.032.